Reliability of non-invasive cardiac output monitoring based on pulse wave transit time in off-pump coronary artery bypass grafting

A. Smetkin
Northern State Medical University, Ministry of Health Care of Russian Federation, 51 Troitskiy Avenue, 163000 Arkhangelsk; City Hospital No. 1 of Arkhangelsk, 1 Suvorova St., 163001 Arkhangelsk
Bio
A. Khusseyn
Northern State Medical University, Ministry of Health Care of Russian Federation, 51 Troitskiy Avenue, 163000 Arkhangelsk
V. Zakharov
Northern State Medical University, Ministry of Health Care of Russian Federation, 51 Troitskiy Avenue, 163000 Arkhangelsk
N. Izotova
Northern State Medical University, Ministry of Health Care of Russian Federation, 51 Troitskiy Avenue, 163000 Arkhangelsk
V. Kuzkov
Northern State Medical University, Ministry of Health Care of Russian Federation, 51 Troitskiy Avenue, 163000 Arkhangelsk; City Hospital No. 1 of Arkhangelsk, 1 Suvorova St., 163001 Arkhangelsk
M. Kirov
Northern State Medical University, Ministry of Health Care of Russian Federation, 51 Troitskiy Avenue, 163000 Arkhangelsk; City Hospital No. 1 of Arkhangelsk, 1 Suvorova St., 163001 Arkhangelsk

Published 2016-08-17

Keywords

  • cardiac output,
  • hemodynamics,
  • pulse wave transit time,
  • transpulmonary thermodilution.

How to Cite

Smetkin, A., Khusseyn, A., Zakharov, V., Izotova, N., Kuzkov, V., & Kirov, M. (2016). Reliability of non-invasive cardiac output monitoring based on pulse wave transit time in off-pump coronary artery bypass grafting. Patologiya Krovoobrashcheniya I Kardiokhirurgiya, 20(2), 104–110. https://doi.org/10.21688/1681-3472-2016-2-104-110

Abstract

Aim: The study was designed to evaluate the accuracy of cardiac output (CO) measurements based on pulse wave transit time (PWTT) as compared to transpulmonary thermodilution (TPTD) in perioperative off-pump coronary artery bypass grafting (OPCABG).
Methods: 20 patients scheduled for elective OPCABG were enrolled for a prospective study. Parallel measurements of CO by using PWTT (COPWTT) and TPTD (COTPTD) were carried out at five stages intraoperatively and at three stages postoperatively. Also analyzed was the agreement between CO absolute values and its ability to monitor cardiac output dynamics.
Results: All in all, 153 pairs of CO data were obtained. According to Bland–Altman analysis, the mean intraoperative difference between the methods was 1.2 L/min, with boundaries of agreement reaching ± 2.9 L/min and a percentage error amounting to 64%. The intraoperative analysis of a polar plot revealed an increased angular difference of 6.9º and radial boundaries of agreement of ± 70.4º, and a decreased polar concordance rate (55%). The postoperative analysis of the data showed better agreement between the methods: mean difference was 0.4 L/min with boundaries of agreement of ± 2.3 L/min and percentage error of 41%. The postoperative trending ability was also more accurate (angular difference of 3.2º, angular boundaries of agreement ± 39º, polar concordance rate 81%).
Conclusion: Poor accuracy and reproducibility of non-invasive COPWTT and its insufficient trending ability to monitor cardiac output during off-pump coronary artery bypass grafting makes it impossible to recommend this technology to be used intraoperatively as an alternative to transpulmonary thermodilution technique. The calibration and trending ability of COPWTT improved after OPCAB.

References

  1. Gosling RG, Budge MM. Terminology for describing the elastic behavior of arteries. Hypertensio 2003;41(6):1180-2.
  2. Ishihara H, Sugo Y, Tsutsui M, Yamada T, Sato T, Akazawa T, Sato N, Yamashita K, Takeda J. The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm. J Clin Monit Comput. 2012;26(6):465-71. http://dx.doi.org/10.1007/s10877-012-9384-7
  3. Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15(2):85-91.
  4. Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25(3):536-46. http://dx.doi.org/10.1053/j.jvca.2011.01.003
  5. Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. 2010;113(5): 1220-35. http://dx.doi.org/10.1097/ALN.0b013e3181ee3130
  6. Sharwood-Smith G, Bruce J, Drummond G. Assessment of pulse transit time to indicate cardiovascular changes during obstetric spinal anaesthesia. Br J Anaesth. 2006;96(1):100-5.
  7. Bataille B, Bertuit M, Mora M, Mazerolles M, Cocquet P, Masson B, Moussot PE, Ginot J, Silva S, Larché J. Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth. 2012;109(6):879-86. http://dx.doi.org/10.1093/bja/aes298
  8. Sinha AC, Singh PM, Grewal N, Aman M, Dubowitz G. Comparison between continuous non-invasive estimated cardiac output by pulse wave transit time and thermodilution method. Ann Card Anaesth. 2014;17(4):273-7. http://dx.doi.org/10.4103/0971-9784.142059
  9. Fischer MO, Balaire X, Le Mauff de Kergal C, Boisselier C, Gérard JL, Hanouz JL, Fellahi JL. The diagnostic accuracy of estimated continuous cardiac output compared with transthoracic echocardiography. Can J Anaesth. 2014;61(1):19-26. http://dx.doi.org/10.1007/s12630-013-0055-z
  10. Sugo Y, Sakai T, Terao M, Ukawa T, Ochiai R. The comparison of a novel continuous cardiac output monitor based on pulse wave transit time and echo Doppler during exercise. Conf Proc IEEE Eng Med Biol So 2012;2012:236-9. http://dx.doi.org/10.1109/EMBC.2012.6345913
  11. Biancofiore G, Critchley LA, Lee A, Bindi L, Bisà M, Esposito M, Meacci L, Mozzo R, DeSimone P, Urbani L, Filipponi F. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. J. Anaesth. 2009;102(1):47-54. http://dx.doi.org/10.1093/bja/aen343
  12. Ishihara H, Tsutsui M. Impact of changes in systemic vascular resistance on a novel non-invasive continuous cardiac output measurement system based on pulse wave transit time: a report of two cases. J Clin Monit Comput. 2014;28(4):423-7. http://dx.doi.org/10.1007/s10877-013-9529-3
  13. Maeda T, Yoshitani K, Inatomi Y, Ohnishi Y. Inaccuracy of the FloTrac/Vigileo™ system in patients with low cardiac index. J Cardiothorac Vasc Anesth. 2014;28(6):1521-6. http://dx.doi.org/10.1053/j.jvca.2014.04.013
  14. Mayer J, Boldt J, Schöllhorn T, Röhm KD, Mengistu AM, Suttner S. Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007;98(2):176-82.
  15. Ribezzo S, Spina E, Di Bartolomeo S, Sanson G. Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU. Scientific World Journal. 2014;2014: http://dx.doi.org/10.1155/2014/353628