Nitrogen dioxide as a by-product of inhaled nitric monoxide therapy. Concentration in the inhaled gas mixture acceptable in clinical practice. A systematic review
Published 2025-04-28
Keywords
- Newborn,
- Nitric Oxide,
- Nitrogen Dioxide,
- Pulmonary Hypertension
How to Cite
Copyright (c) 2025 Bautin A.E., Chomakhashvili I.G., Afanasyeva K.Yu., Marichev A.O., Radovskiy A.M., Kozyrev D.A., Barygin E.K., Ablesimov G.N., Badalyan N.V., Zhuravleva M.S., Pozhidaeva A.M., Burov A.A., Valueva Yu.V.

This work is licensed under a Creative Commons Attribution 4.0 International License.
Abstract
Objective: The study aimed to review the literature data on the recommended maximum concentrations of nitrogen dioxide during inhaled therapy with nitric oxide and to propose a general list of recommendations on safe therapy with inhaled nitric oxide.
Methods: The systematic review was performed according to the PRISMA 2020 criteria. A search was carried out in databases Russian Science Citation Index, MedLine, Cochrane Library using the keywords “nitrogen dioxide”, “limit concentration”, “annual limit value”, “inhaled nitric oxide”.
Results: The review included 28 articles, 20 sources from regulatory authorities setting hygiene standards, and 2 instructions for devices.
Conclusion: For short periods of NO inhalation not exceeding one hour, the maximum permissible concentration of NO2 in the inhaled gas mixture may be defined as 5 ppm. In cases of NO inhalation not exceeding 24 hours, the maximum permissible concentration of NO2 in the inhaled mixture should be considered as 2 ppm. If the duration of NO inhalation exceeds 24 hours, the permissible concentration of NO2 in the inhaled gas mixture should be limited to 1 ppm.
Received 29 November 2024. Revised 9 January 2025. Accepted 13 January 2025.
Funding
This research was carried out within the state assignment “Development of a new device for supplying nitric oxide synthesized from atmospheric air into mechanical circulatory assist devices” No. 123021000129‑1.
Conflict of interest
The authors declare no conflict of interest.
Contribution of the authors
Conception and study design: A.E. Bautin, A.A. Burov, Yu.V. Valueva
Data collection and analysis: I.G. Chomakhashvili, K.Yu. Afanasyeva, D.A. Kozyrev, E.K. Barygin, G.N. Ablesimov, N.V. Badalyan, M.S. Zhuravleva, A.M. Pozhidaeva, A.A. Burov, Yu.V. Valueva
Drafting the article: I.G. Chomakhashvili, A.A. Burov, Yu.V. Valueva
Critical revision of the article: A.E. Bautin, A.O. Marichev, A.M. Radovskiy, A.A. Burov
Final approval of the version to be published: A.E. Bautin, I.G. Chomakhashvili, K.Yu. Afanasyeva, A.O. Marichev, A.M. Radovskiy, D.A. Kozyrev, E.K. Barygin, G.N. Ablesimov, N.V. Badalyan, M.S. Zhuravleva, A.M. Pozhidaeva, A.A. Burov, Yu.V. Valueva
References
- Ghadimi K., Cappiello J.L., Wright M.C., Levy J.H., Bryner B.S., DeVore A.D., Schroder J.N., Patel C.B., Rajagopal S., Shah S.H., Milano C.A.; INSPIRE-FLO Investigators. Inhaled epoprostenol compared with nitric oxide for right ventricular support after major cardiac surgery. Circulation. 2023;148(17):1316-1329. PMID: 37401479; PMCID: PMC10615678. https://doi.org/10.1161/CIRCULATIONAHA.122.062464
- Redaelli S., Pozzi M., Giani M., Magliocca A., Fumagalli R., Foti G., Berra L., Rezoagli E. Inhaled nitric oxide in acute respiratory distress syndrome subsets: rationale and clinical applications. J Aerosol Med Pulm Drug Deliv. 2023;36(3):112-126. PMID: 37083488; PMCID: PMC10402704. https://doi.org/10.1089/jamp.2022.0058
- Faustini A., Rapp R., Forastiere F. Nitrogen dioxide and mortality: review and meta-analysis of long-term studies. Eur Respir J. 2014;44(3):744-753. PMID: 24558178. https://doi.org/10.1183/09031936.00114713
- WHO global air quality guidelines. Particulate matter (PM 2.5 and PM 10), ozone, nitrogen dioxide, sulfur dioxide and carbon monoxide. Geneva: World Health Organization Publ.; 2021. Pp. 111-125.
- da Rocha F.O.C., Barbosa J.L., Campos V.P., Alves R.S. Nitrogen oxides levels in the atmosphere of different Brazilian urban centers, by passive sampling. J Braz Chem Soc. 2022;33(2):143-156. https://doi.org/10.21577/0103-5053.20210131
- ГОСТ 12.1.005–88. Межгосударственный стандарт. Система стандартов безопасности труда. Общие санитарно-гигиенические требования к воздуху рабочей зоны: утв. и введен в действие Постановлением Госстандарта СССР от 29.09.1988 № 3388 (ред. от 20.06.2000). GOST 12.1.005–88. Interstate standard. Occupational safety standards system. General sanitary and hygienic requirements for working zone air. Approved and put into effect by Resolution of the USSR State Standard of 29.09.1988 no. 3388 (as amended on 20.06.2000). (In Russ.)
- Предельно допустимые концентрации загрязняющих веществ в атмосферном воздухе городских и сельских поселений: гигиенические нормативы ГН 2.1.6.3492-17 (утв. постановлением Главного государственного санитарного врача Российской Федерации 31.05.2018 № 37). М.: Федеральный центр гигиены и эпидемиологии Роспотребнадзора, 2019. 55 с. Maximum permissible concentrations of pollutants in the atmospheric air of urban and rural settlements: hygienic standards ГН 2.1.6.3492-17 (approved by the resolution of the Chief State Sanitary Doctor of the Russian Federation of 31.05.2018 no. 37). Moscow: Federal Center for Hygiene and Epidemiology of Federal Service for the Oversight of Consumer Protection and Welfare Publ.; 2019. 55 p. (In Russ.)
- Carson T.R., Rosenholtz M.S., Wilinski F.T., Weeks M.H. The responses of animals inhaling nitrogen dioxide for single, short-term exposures. Am Ind Hyg Assoc J. 1962;23:457-462. PMID: 14018928. https://doi.org/10.1080/00028896209342900
- Hine C.H., Meyers F.H., Wright R.W., Dewey M.L. Pulmonary changes in animals exposed to nitrogen dioxide. Proc West Pharmacol Soc. 1964;7:19-22. PMID: 14185232.
- Hugod C. Effect of exposure to 43 ppm nitric oxide and 3.6 ppm nitrogen dioxide on rabbit lung. A light and electron microscopic study. Int Arch Occup Environ Health. 1979;42(3-4):159-167. PMID: 422258. https://doi.org/10.1007/BF00377770
- Goldstein E., Peek N.F., Parks N.J., Hines H.H., Steffey E.P., Tarkington B. Fate and distribution of inhaled nitrogen dioxide in rhesus monkeys. Am Rev Respir Dis. 1977;115(3):403-412. PMID: 402875. https://doi.org/10.1164/arrd.1977.115.3.403
- Han M., Guo Z., Li G., Sang N. Nitrogen dioxide inhalation induces genotoxicity in rats. Chemosphere. 2013;90(11):2737-2742. PMID: 23332788. https://doi.org/10.1016/j.chemosphere.2012.11.057
- Hesterberg T.W., Bunn W.B., McClellan R.O., Hamade A.K., Long C.M., Valberg P.A. Critical review of the human data on short-term nitrogen dioxide (NO2) exposures: evidence for NO2 no-effect levels. Crit Rev Toxicol. 2009;39(9):743-781. PMID: 19852560. https://doi.org/10.3109/10408440903294945
- Barck C., Sandström T., Lundahl J., Halldén G., Svartengren M., Strand V., Rak S., Bylin G. Ambient level of NO2 augments the inflammatory response to inhaled allergen in asthmatics. Respir Med. 2002;96(11):907-917. PMID: 12418589. https://doi.org/10.1053/rmed.2002.1374
- Ahmed T., Dougherty R., Sackner M.A. Effect of NO2 exposure on specific bronchial reactivity in subjects with allergic bronchial asthma [final report]. Contract report no. CR-83/07/BI. Warren: General Motors Research Laboratories, 1983.
- Kleinman M.T., Bailey R.M., Linn W.S., Anderson K.R., Whynot J.D., Shamoo D.A., Hackney J.D. Effects of 0.2 ppm nitrogen dioxide on pulmonary function and response to bronchoprovocation in asthmatics. J Toxicol Environ Health. 1983;12(4-6):815-826. PMID: 6668625. https://doi.org/10.1080/15287398309530472
- Phillips M.L., Hall T.A., Sekar K., Tomey J.L. Assessment of medical personnel exposure to nitrogen oxides during inhaled nitric oxide treatment of neonatal and pediatric patients. Pediatrics. 1999;104(5 Pt 1):1095-1100. PMID: 10545553. https://doi.org/10.1542/peds.104.5.1095
- Foubert L., Fleming B., Latimer R., Jonas M., Oduro A., Borland C., Higenbottam T. Safety guidelines for use of nitric oxide. Lancet. 1992;339(8809):1615-1616. PMID: 1351591. https://doi.org/10.1016/0140 6736(92)91886-d
- Frostell C.G., Zapol W.M. Inhaled nitric oxide, clinical rationale and applications. Adv Pharmacol. 1995;34:439-456. PMID: 8562450. https://doi.org/10.1016/s1054-3589(08)61102-3
- Davidson D., Barefield E.S., Kattwinkel J., Dudell G., Damask M., Straube R., Rhines J., Chang C.T. Inhaled nitric oxide for the early treatment of persistent pulmonary hypertension of the term newborn: a randomized, double-masked, placebo-controlled, dose-response, multicenter study. The I-NO/PPHN Study Group. Pediatrics. 1998;101(3 Pt 1):325-334. PMID: 9480993. https://doi.org/10.1542/peds.101.3.325
- Miller C., Miller M., McMullin B., Regev G., Serghides L., Kain K., Road J., Av-Gay Y. A phase I clinical study of inhaled nitric oxide in healthy adults. J Cyst Fibros. 2012;11(4):324-331. PMID: 22520076. https://doi.org/10.1016/j.jcf.2012.01.003
- Goldbart A., Golan-Tripto I., Pillar G., Livnat-Levanon G., Efrati O., Spiegel R., Lubetzky R., Lavie M., Carmon L., Ghaffari A., Nahum A. Inhaled nitric oxide therapy in acute bronchiolitis: a multicenter randomized clinical trial. Sci Rep. 2020;10(1):9605. Erratum in: Sci Rep. 2020;10(1):17640. PMID: 32541773; PMCID: PMC7295966. https://doi.org/10.1038/s41598-020-66433-8
- Wiegand S.B., Safaee Fakhr B., Carroll R.W., Zapol W.M., Kacmarek R.M., Berra L. Rescue treatment with high-dose gaseous nitric oxide in spontaneously breathing patients with severe coronavirus disease 2019. Crit Care Explor. 2020;2(11):e0277. PMID: 33225304; PMCID: PMC7671879. https://doi.org/10.1097/CCE.0000000000000277
- Safaee Fakhr B., Di Fenza R., Gianni S., Wiegand S.B., Miyazaki Y., Araujo Morais C.C., Gibson L.E., Chang M.G., Mueller A.L., Rodriguez-Lopez J.M., Ackman J.B., Arora P., Scott L.K., Bloch D.B., Zapol W.M., Carroll R.W., Ichinose F., Berra L.; Nitric Oxide Study Investigators. Inhaled high dose nitric oxide is a safe and effective respiratory treatment in spontaneous breathing hospitalized patients with COVID-19 pneumonia. Nitric Oxide. 2021;116:7-13. PMID: 34400339; PMCID: PMC8361002. https://doi.org/10.1016/j.niox.2021.08.003
- Fung C., Z'Graggen W.J., Jakob S.M., Gralla J., Haenggi M., Rothen H.-U., Mordasini P., Lensch M., Söll N., Terpolilli N., Feiler S., Oertel M.F., Raabe A., Plesnila N., Takala J., Beck J. Inhaled nitric oxide treatment for aneurysmal SAH patients with delayed cerebral ischemia. Front Neurol. 2022;13:817072. PMID: 35250821; PMCID: PMC8894247. https://doi.org/10.3389/fneur.2022.817072
- Valsecchi C., Winterton D., Safaee Fakhr B., Collier A.-R.Y., Nozari A., Ortoleva J., Mukerji S., Gibson L.E., Carroll R.W., Shaefi S., Pinciroli R., La Vita C., Ackman J.B., Hohmann E., Arora P., Barth W.H. Jr, Kaimal A., Ichinose F., Berra L.; DELiverly oF iNO (DELFiNO) Network Collaborators. High-dose inhaled nitric oxide for the treatment of spontaneously breathing pregnant patients with severe coronavirus disease 2019 (COVID-19) pneumonia. Obstet Gynecol. 2022;140(2):195-203. PMID: 35852269; PMCID: PMC9994457. https://doi.org/10.1097/AOG.0000000000004847
- Di Fenza R., Shetty N.S., Gianni S., Parcha V., Giammatteo V., Safaee Fakhr B., Tornberg D., Wall O., Harbut P., Lai P.S., Li J.Z., Paganoni S., Cenci S., Mueller A.L., Houle T.T., Akeju O., Bittner E.A., Bose S., Scott L.K., Carroll R.W., Ichinose F., Hedenstierna M., Arora P., Berra L.; Nitric Oxide Investigators. High-dose inhaled nitric oxide in acute hypoxemic respiratory failure due to COVID-19: a multicenter phase II trial. Am J Respir Crit Care Med. 2023;208(12):1293-1304. PMID: 37774011; PMCID: PMC10765403. https://doi.org/10.1164/rccm.202304-0637OC
- Matsugi E., Takashima S., Doteguchi S., Kobayashi T., Okayasu M. Real-world safety and effectiveness of inhaled nitric oxide therapy for pulmonary hypertension during the perioperative period of cardiac surgery: a post-marketing study of 2 817 patients in Japan. Gen Thorac Cardiovasc Surg. 2024;72(5):311-323. PMID: 37713058; PMCID: PMC11018662. https://doi.org/10.1007/s11748-023-01971-2
- Кравченко И.В., Подоксенов Ю.К., Тё М.А., Калашникова Т.П., Чурилина Е.А., Козулин М.С., Гусакова А.М., Козлов Б.Н., Каменщиков Н.О. Профилактика дыхательных осложнений у пациентов с риском неблагоприятных респираторных событий путем периоперационной доставки оксида азота при кардиохирургических операциях в условиях искусственного кровообращения: одноцентровое проспективное рандомизированное исследование. Патология кровообращения и кардиохирургия. 2024;28(3):78-93. https://doi.org/10.21688/1681-3472-2024-3-78-93 Kravchenko I.V., Podoksenov Yu.K., Tyo M.A., Kalashnikova T.P., Churilina E.A., Kozulin M.S., Gusakova A.M., Kozlov B.N., Kamenshchikov N.O. Prevention of respiratory complications in patients at risk of adverse respiratory events by perioperative nitric oxide delivery during cardiac surgery under cardiopulmonary bypass: a single-center prospective randomized study. Patologiya krovoobrashcheniya i kardiokhirurgiya = Circulation Pathology and Cardiac Surgery. 2024;28(3):78-93. (In Russ.) https://doi.org/10.21688/1681-3472-2024-3-78-93
- Bouchet M., Renaudin M.H., Raveau C., Mercier J.C., Dehan M., Zupan V. Safety requirement for use of inhaled nitric oxide in neonates. Lancet. 1993;341(8850):968-969. PMID: 8096310. https://doi.org/10.1016/0140-6736(93)91267-p
- Miller O.I., Celermajer D.S., Deanfield J.E., Macrae D.J. Guidelines for the safe administration of inhaled nitric oxide. Arch Dis Child Fetal Neonatal Ed. 1994;70(1):F47-F49. PMID: 8117128; PMCID: PMC1060988. https://doi.org/10.1136/fn.70.1.f47
- Miyamoto K., Aida A., Nishimura M., Nakano T., Kawakami Y., Ohmori Y., Ando S., Ichida T. Effects of humidity and temperature on nitrogen dioxide formation from nitric oxide. Lancet. 1994;343(8905):1099-1100. PMID: 7909115. https://doi.org/10.1016/s0140-6736(94)90210-0
- Sokol G.M., Van Meurs K.P., Wright L.L., Rivera O., Thorn W.J., Chu P.M., Sams R.L. Nitrogen dioxide formation during inhaled nitric oxide therapy. Clin Chem. 1999;45(3):382-387. https://doi.org/10.1093/clinchem/45.3.382
- Баутин А.Е., Афанасьева К.Ю., Маричев А.О., Радовский А.М., Барыгин Е.К., Гусенова А.К., Чомахашвили И.Г., Бадалян Н.В., Журавлева М.С., Пожидаева А.М. Влияние состава подаваемой газовой смеси на продукцию диоксида азота в модели ингаляционной терапии оксидом азота (экспериментальное исследование). Анестезиология и реаниматология. 2024;(1):23-30. https://doi.org/10.17116/anaesthesiology202401123 Bautin A.E., Afanasyeva K.Yu., Marichev A.O., Radovskiy A.M., Barygin E.K., Gusenova A.K., Chomakhashvili I.G., Badalyan N.V., Zhuravleva M.S., Pozhidaeva A.M. Impact of inhaled gas compound on nitric dioxide release in a model of nitric oxide inhalation therapy (experimental study). Russian Journal of Anaesthesiology and Reanimatology. 2024;(1):23-30. (In Russ., In Engl.) https://doi.org/10.17116/anaesthesiology202401123
- Liu Y., Zhu Y., Jiang C., Su Z., Yan Y., Feng B., Mao W., Zhang Y., Wang X., Xu Z., Zhang H. An electrochemical nitric oxide generator for in-home inhalation therapy in pulmonary artery hypertension. BMC Med. 2022;20(1):481. PMID: 36522645; PMCID: PMC9753075. https://doi.org/10.1186/s12916-022-02686-6
- Gianni S., Carroll R.W., Kacmarek R.M., Berra L. Inhaled nitric oxide delivery systems for mechanically ventilated and nonintubated patients: a review. Respir Care. 2021;66(6):1021-1028. PMID: 34039761; PMCID: PMC10408459. https://doi.org/10.4187/respcare.08856
- Petit P.C., Fine D.H., Vásquez G.B., Gamero L., Slaughter M.S., Dasse K.A. The pathophysiology of nitrogen dioxide during inhaled nitric oxide therapy. ASAIO J. 2017;63(1):7-13. PMID: 27556146. https://doi.org/10.1097/MAT.0000000000000425